Description
Measurement Time | Approx. 3.5 second (SCI+SCE measurement) |
Approx. 4 second (SCI+SCE+Gloss measurement) | |
Approx. 2.5 second (Transmittance) | |
Approx. 3 second (UV-cut/UV-adjusted; SCI or SCE) | |
Minimum Interval Between Measurements | Approx. 4 second (SCI+SCE measurement) |
Approx. 4.5 second (SCI+SCE+Gloss measurement) | |
Approx. 3 second (Transmittance) | |
Approx. 4 second (UV-cut/UV-adjusted; SCI or SCE) | |
Sample Viewer Function | Using internal camera |
(Image viewable/copiable using optional software such as SpectraMagic NX software Ver. 3.2 or later) | |
Internal Performance Check*2 | WAA (Wavelength Analysis & Adjustment) Technology |
Interface | USB 2.0 |
Target Mask Auto Detection | Yes |
Power | Dedicated AC adapter |
Operating temperature / humidity range | 13 to 33°C, Relative humidity: 80% or less (at 35°C) with no condensation |
Storage temperature / humidity range | 0 to 40°C, Relative humidity: 80% or less (at 35°C) with no condensation |
Size (W x H x D) | Approx. 300 x 677 x 315 mm |
Weight | Approx. 14.0 kg |
Color
Illumination/ Viewing System | Reflectance: |
di:8°, de:8° (diffused illumination, 8° viewing), SCI (Specular Component Included) / SCE (Specular Component Excluded) switchable. | |
Conforms to CIE No.15, lSO7724/1, ASTM E1164, DIN 5033 Teil7, and JIS Z 8722 condition c standards | |
Transmittance: | |
di:0°, de:0° (diffused illumination, 0° viewing) | |
Conforms to CIE No.15, ASTM E1164, DIN 5033 Teil7, and JIS Z 8722 condition g standards | |
Integrating Sphere Size | Ø152 mm (6 inches) |
Detector | Dual 40-element silicon photodiode arrays |
Spectral Separation Device | Diffraction grating |
Wavelength Range | 360 to 740 nm |
Wavelength Pitch | 10 nm |
Half Bandwidth | approx. 10 nm |
Reflectance | 0 to 200% |
Range | Resolution: 0.01% |
Light Source | 3 Pulsed xenon lamps (2 with UV cut filters) |
Illumination / Measurement Area | LAV: Ø30.0 mm / Ø25.4 mm |
LMAV: Ø20.0 mm / Ø16.0 mm | |
MAV: Ø11.0 mm / Ø8.0 mm | |
SAV: Ø7.0 mm / Ø4.0 mm | |
Trans: Ø24.0 mm / Ø17.0 mm | |
Repeatability | Colorimetric values: Standard deviation within ΔE*ab 0.02 |
Spectral reflectance: Standard deviation within 0.1% | |
(When a white calibration plate is measured 30 times at 10-second intervals after white calibration) | |
Inter-Instrument Agreement | Within ΔE*ab 0.12 |
(Based on average for 12 BCRA Series II color tiles; LAV/SCI. Compared to values measured with master body under Konica Minolta standard measurement conditions) | |
UV Setting | 100% / 0% / Adjusted |
Instantaneous numerical adjustment of UV with no mechanical filter movement required*1; 400 nm and 420 nm UV cutoff filters |
Gloss
Measurement Angle | 60° |
Light Source | White LED |
Detector | Silicon photodiode |
Measurement Range | 0 to 200 GU |
Resolution: 0.01 GU | |
Measurement Area | MAV (LAV/LMAV/MAV color measurement area): |
Ø10.0 x 8.0 mm ellipse | |
SAV (SAV color measurement): | |
Ø3.0 mm | |
Repeatability | 0 to 10 GU: within 0.1 GU |
10 to 100 GU: within 0.2 GU | |
100 to 200 GU: 0.2% | |
(When measured 30 times at 10-second intervals) | |
Inter-Instrument Agreement | 0 to 10 GU: ±0.2 GU |
10 to 100 GU: ±0.5 GU | |
(MAV. Compared to values measured with master body under Konica Minolta standard conditions) | |
Geometry | JIS Z 8741 (MAV), JIS K 5600, ISO 2813, ISO 7668 (MAV), ASTM D523-08, ASTM D2457-13, and DIN 67530 |